FLUID SHEAR STEM CELL BIOREACTOR DAVID REETZ, GAGE NARDI, GUINEVERE RICHMOND, NICHOLAS ANDERSON & RUI BAO

Design, build, and evaluate a low-cost bioreactor system to apply user-defined magnitudes of shear stress to mammalian cells in culture to explore scientific questions in mechanobiology

Flow Path in Bioreactor

BACKGROUND & VALUE PROP Cells embedded in tendons experience shear stress as collagen fibers slide past each other Dr. Nathan Schiele's lab is interested in studying how shear stresses on stem cells impact tendon lineage and tendon formation Aim to create a bioreactor that exposes stem cells to a range of shear stresses that represents mechanical stimuli

System Schematic Design

Pump Performance Curve

FINAL DESIGN

We aim to create a low-cost bioreactor for simulating in vivo stress conditions of tenocytes in cell culture with a userfriendly interface that biologists can use without technical knowledge of the reactor itself.

KEY REQUIREMENTS Withstand incubator conditions for cell cultures ▪ Fit inside a 14” x 12” x 12” space ▪ 37 °C, 5 % CO2, & 95% Humidity Apply long term shear stress at 0-150 mPa Apply short term shear stress at 1-2 Pa Low-cost & user friendly Components must be autoclavable or ethanol friendly Accommodate cell culture plates with minimum cell culture usage

Valve, stepper motor and sensor used, full view of resin printed bioreactor, and top view of bioreactor and reservoir unit

Criteria Met Two flow paths with different valves and sensors to meet both shear stress ranges specified Low total cost even with specialized parts Utilization of off-the-shelf products where applicable Easy to use GUI interface All parts are autoclavable Reservoir size and subsequent media usage is minimal Withstands incubator conditions Fluid flow is completely contained

VALIDATION Analytical Modeling ▪ Using Poiseuille flow, a desired flow rate was found for the bioreactor’s channel Flow Testing ▪ Compared the sensor reading to a manual reading of flow rate to confirm accuracy Leak Testing Hardware Integration

Low Flow Sensor Accuracy Sensor Reading (ml/min)

CONCEPT DEVELOPMENT

45

y = 0.9607x - 1.3865 R² = 0.9928

40 35 30 25 20 15 10 5 0 0

5

10

15

20

25

30

35

40

45

50

Manual Reading (ml/min)

High Flow Sensor Accuracy Sensor Reading (ml/min)

OBJECTIVE

700 600

y = 1.0124x - 6.3004 R² = 0.9998

500 400 300 200 100 0 0

100

200

300

400

500

600

700

Manual Reading (ml/min)

CONCLUSION The bioreactor system can deliver fluid shear stresses by controlling the flow rate within the bioreactor system and is able to apply user-defined shear stress(s) for a specified time. The system is operated independently by a Raspberry Pi and Arduino. In future work, engineers should focus on making a simpler priming process, downscale the size of the footprint of the reservoir, and speed up the time it takes for the valves to be adjusted correctly. More designing is needed to integrate another bioreactor with the current system and increase the number of different experiments the system can run.

ACKNOWLEDGEMENTS This project was funded by Dr. Nathan Schiele for his tendon tissue engineering lab at the University of Idaho. Special thanks to our instructor Dr. Russel Qualls and our mentor Colin Marchus.