Osstell Beacon | How to measure 1
Activate the Osstell Beacon by picking it up then place a barrier sleeve over the instrument. The instrument will start-up and be ready for measurement in the BL (Buccal – Lingual) direction.
2
Attach the SmartPeg to the implant or abutment by screwing the SmartPeg Mount using finger force of approximately 4-6 Ncm.
Hold the instrument tip close (3-5mm) to the top of the SmartPeg without touching it. A sound indicates when measurement has started and the value will be shown in the upper display combined with a colored light indication below the instrument tip.
3
Bring the instrument out of the mouth to clearly read the ISQ value and the colored indication. The measured ISQ values will be displayed in the upper display for a couple of seconds and then switch to indicate ready for measurement in mesial-distal direction.
BL MD